POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter with open suture holes, REF...
Bard Peripheral Vascular Inc began a recall of POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter with open suture holes, REF 5678301 on . FDA classified it as Class II and gives the reason as "The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1771-2018
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter with open suture holes, REF 5678301
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
- Quantity in commerce
- 181 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Number Unique Device Identifier (UDI) REAN0061 (01)00801741110818(17)170628(10)REAN0061 REAP0772 (01)00801741110818(17)170728(10)REAP0772 REAT1524 (01)00801741110818(17)171128(10)REAT1524 REAV1479 (01)00801741110818(17)171228(10)REAV1479 REAV1938 (01)00801741110818(17)171028(10)REAV1938 REAW0937 (01)00801741110818(17)171031(10)REAW0937 REAW1186 (01)00801741110818(17)171031(10)REAW1186 REAW1666 (01)00801741110818(17)171031(10)REAW1666 REAY0528 (01)00801741110818(17)171031(10)REAY0528 REAY0795 (01)00801741110818(17)171031(10)REAY0795 REAZ1529 (01)00801741110818(17)171231(10)REAZ1529 REBN0024 (01)00801741110818(17)171231(10)REBN0024 REBN1581 (01)00801741110818(17)180131(10)REBN1581 REBQ0210 (01)00801741110818(17)180131(10)REBQ0210 REBS1277 (01)00801741110818(17)181031(10)REBS1277 REBR1125 (01)00801741110818(17)181031(10)REBR1125 REBS1395 (01)00801741110818(17)181031(10)REBS1395 REBS2412 (01)00801741110818(17)181130(10)REBS2412 REBT1785 (01)00801741110818(17)181231(10)REBT1785 REBT2349 (01)00801741110818(17)181231(10)REBT2349 REBU0563 (01)00801741110818(17)181231(10)REBU0563 REBV0760 (01)00801741110818(17)190228(10)REBV0760 REBW1037 (01)00801741110818(17)190331(10)REBW1037
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1760-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1761-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1762-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1763-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1764-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1765-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1766-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1767-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1768-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1769-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1770-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1771-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1772-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1773-2018 | Class II | Completed | Voluntary: Firm initiated |