510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1759-2018Class II

Retractor f/Sciatic Nerve Long; Part number: 03.100.014; UDI: 10886982070630

Synthes (USA) Products LLC / initiated 2018-04-02 / Terminated

Synthes (USA) Products LLC began a recall of Retractor f/Sciatic Nerve Long; Part number: 03.100.014; UDI: 10886982070630 on . FDA classified it as Class II and gives the reason as "There is a potential for microspores to form on the hollow handle of the Sciatic Nerve Retractor. The pores may increase in size, allowing fluid to enter the hollow handle". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1759-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Retractor f/Sciatic Nerve Long; Part number: 03.100.014; UDI: 10886982070630
Product code
GAD Retractor
Reason for recall
There is a potential for microspores to form on the hollow handle of the Sciatic Nerve Retractor. The pores may increase in size, allowing fluid to enter the hollow handle.
FDA root cause
Under Investigation by firm
Action
On April 2, 2018 Synthes (USA) LLC mailed an Urgent Product Recall notification letter to affected customers. Customers were instructed to: Review current inventory to identify and quarantine all affected products; Complete the Verification section of the Return Receipt form; Return affected product; Forward the notification to anyone in your facility that needs to be informed; If affected products have been forwarded to another facility, contact that facility to arrange return; Keep a copy of the notification. Customers with questions may call (610)719-5450 or contact your Depuy Synthese Sales Consultant.
Quantity in commerce
196
Distribution
US Nationwide and Canada
Product codes and lots
Lot numbers: T104993, T108115, T114598, T140566, T140665, T140670, T140671, T140673, T141540, T143687, T144854, T145585, T145590, T145932, T146657, T147929, T148552, T148553, T149727, T149728 T151369, T151488, T152808, T152809, T153884, T155649, T958061

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1759-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1758-2018Class IITerminatedVoluntary: Firm initiated
Z-1759-2018Class IITerminatedVoluntary: Firm initiated