Z-1893-2018Class II

iChem VELOCITY Urine Chemistry Strips, REF 800-7212

Siemens Healthcare Diagnostics Inc. / initiated 2017-12-04 / Terminated

Beckman Coulter Inc. began a recall of iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to... on . FDA classified it as Class II and gives the reason as "A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1893-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChem VELOCITY strips are intended for use with the iCHEM VELOCITY analyzer. In particular they are not intended for visual reading. The iChem VELOCITY test strips are not intended for visual reading. The iChem VELOCITY is not intended to be used as a Point of Care (POC) analyzer. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iCHEM VELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.
Product code
LJX Test, Urine Leukocyte
Reason for recall
A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.
FDA root cause
Mixed-up of materials/components
Action
The firm initiated the recall by letter on 12/04/2017. The letter identified the affected product, problem and the actions to be taken. The firm requested destruction of the recalled product.
Quantity in commerce
14340 units
Distribution
Worldwide Distribution and US Nationwide Distribution
Product codes and lots
Lot #7212154 A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1893-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1893-2018Class IITerminatedVoluntary: Firm initiated