Syngo.Via VB20A model 1049610
Siemens Medical Solutions USA, Inc began a recall of Syngo.Via VB20A model 1049610 on . FDA classified it as Class II and gives the reason as "Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1778-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo.Via VB20A model 1049610
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 03/27/18 Firm became aware that recall was necessary Firm initiated recall on 03/27/18 via customer information letter SY007/18/P advising that corrected values modified in the report are not save. Software patch VB20A_HF05 is being released to resolve the matter
- Distribution
- Nationwide
- Product codes and lots
- All
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1778-2018 | Class II | Terminated | Voluntary: Firm initiated |