Z-1875-2018Class II
Epic Extremity Plate System,10x10 Plate Caddy, Part 2111-1010 with 2.7 and 3.5mm Non-Locking Screw...
Epic Extremity, LLC / initiated 2018-03-20 / Terminated / root cause: labeling and UDI
Epic Extremity, LLC began a recall of Epic Extremity Plate System,10x10 Plate Caddy, Part 2111-1010 with 2.7 and 3.5mm Non-Locking Screw (2001-27XX-N & 2001-35XX-N) on . FDA classified it as Class II and gives the reason as "There is an error in the caddy artwork that misidentifies a specific plate and screw". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1875-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Epic Extremity, LLC
- Product
- Epic Extremity Plate System,10x10 Plate Caddy, Part 2111-1010 with 2.7 and 3.5mm Non-Locking Screw (2001-27XX-N & 2001-35XX-N)
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- There is an error in the caddy artwork that misidentifies a specific plate and screw.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Notification letters were sent on 3/20/18. The letters instructed customers to: We ask you to be aware of the incorrect numbers and ensure you are ordering the correct part numbers until replacement parts are available. You will be informed promptly of the availability of the replacement trays. You will be contacted to arrange return and to exchange the parts presently in your possession. Our records indicate that you have received one or more of the trays involved. Therefore, we would ask you to reply to this memo and confirm the quantities of each tray in your possession. Your response will serve as acknowledgement of this notification and that you will act in compliance. Please reply to [email protected]. The notification letter included photographs of products to indicate errors in labeling.
- Quantity in commerce
- 9 units
- Distribution
- The products were distributed to the following US states: FL, PA, and VA.
- Product codes and lots
- Catalog Number: 2111-1010 Lot: 51461
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1874-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1875-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.