Bard Peripheral Vascular Inc began a recall of BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature on . FDA classified it as Class II and gives the reason as "Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1802-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emailed to [email protected] or faxed to 1-800-994-6772.
- Quantity in commerce
- 140 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- UDI: (01)00801741067921(17)191115(10)VTBY0563, Lot Number: VTBY0563
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1802-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1803-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.