Z-1803-2018Class II

BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.014", 150 cm, REF SK15014. Intended to...

Bard Peripheral Vascular Inc / initiated 2018-04-16 / Terminated / root cause: manufacturing process

Bard Peripheral Vascular Inc began a recall of BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.014", 150 cm, REF SK15014. Intended to support a guidewire during access into the vasculature on . FDA classified it as Class II and gives the reason as "Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1803-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.014", 150 cm, REF SK15014. Intended to support a guidewire during access into the vasculature.
Product code
DQY Catheter, Percutaneous
Reason for recall
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emailed to [email protected] or faxed to 1-800-994-6772.
Quantity in commerce
130 units
Distribution
Nationwide Distribution
Product codes and lots
UDI: (01)00801741067907(17)191207(10)VTBZ0193, Lot Number: VTBZ0193

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1803-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1802-2018Class IITerminatedVoluntary: Firm initiated
Z-1803-2018Class IITerminatedVoluntary: Firm initiated