The BD Vacutainer Plus Plastic Citrate Blood Collection Tube is an evacuated blood collection tube that...
Becton Dickinson & Company began a recall of The BD Vacutainer Plus Plastic Citrate Blood Collection Tube is an evacuated blood collection tube that provides a means of collecting, transporting, and... on . FDA classified it as Class II and gives the reason as "Approximately 3.4% of BD Vacutainer Plus Plastic Citrate Blood Collection Tubes (2.7mL) (light blue top tube) will exhibit over fill draw volumes from ~11-14%". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2258-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- The BD Vacutainer Plus Plastic Citrate Blood Collection Tube is an evacuated blood collection tube that provides a means of collecting, transporting, and processing blood in a closed tube. The buffered sodium citrate additive provides an anticoagulated specimen that may be used for clinical laboratory coagulation assays.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- Approximately 3.4% of BD Vacutainer Plus Plastic Citrate Blood Collection Tubes (2.7mL) (light blue top tube) will exhibit over fill draw volumes from ~11-14%.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- BD began notifying distributors on April 11, 2018. Customer communications were mailed on April 12-13th, 2018. US customers will be sent a letter via UPS beginning April 12, 2018. Distributors were e-mailed on April 11, 2018 and requested that they provide a list of their consignees to facilitate mailing to end users BD Regions will be contacted via email regarding distribution outside of the US Distributors were instructed to return customer lists to BD. Letters will be mailed with UPS tracking, which provides for a 100% effectiveness check. Customer response form will be tabulated. 100% of non-responders will be contacted by phone.
- Distribution
- US/PR-All 50 states and PR Global-AR, BO, BR, BS, BZ, CA, CO, CR, CW, DO, EC, GT, GY, IN, JM, JP, KR, MX, NI, PA, PE, SG, SR, SV, TT, TW, UY, VI Dates of distribution: June 2, 2017 to the present.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2258-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2259-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2260-2018 | Class II | Terminated | Voluntary: Firm initiated |