Jasper Vektor, Class II Correction Appliance, Part No. 610-536R used for correcting class 2 malocclusions by...
TP Orthodontics, Inc. began a recall of Jasper Vektor, Class II Correction Appliance, Part No. 610-536R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube... on . FDA classified it as Class II and gives the reason as "The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1795-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TP Orthodontics, Inc.
- Product
- Jasper Vektor, Class II Correction Appliance, Part No. 610-536R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.
- Product code
- ECO Spring, Orthodontic
- Reason for recall
- The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On May 3, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter advised customers that the firm was voluntarily recalling Generation 1 Jasper Vektor Class 2 Correction Appliances due to receiving 106 customer complaints that the device had broken in the patient's mouth during placement and usage. Customers were instructed to do the following: 1. Locate any of the affected lots identified in the table below and quarantine the product. 2. Return the Generation 1 Jasper Vektor for a replacement Jasper Vektor Generation 2 or refund, following the instructions below: Contact TP Orthodontics, Inc. at 800-338-8856 or [email protected] for a Return Goods Authorization and pre-paid shipping label. Alternately, use the carrier of your choice with available tracking and send to: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 3. Email, mail, or fax the attached response letter to TP Orthodontics, Inc. In addition to sending the letter back, you may call us with any questions or concerns. EMAIL: [email protected] FAX: 219-324-3029 MAIL: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 4. If you have any further questions or concerns after reviewing this notice, please contact TPO Customer Care Team 800-388-8856 or [email protected]
- Quantity in commerce
- 1
- Distribution
- Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.
- Product codes and lots
- 25915000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1783-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1784-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1785-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1786-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1787-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1788-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1789-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1790-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1791-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1792-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1793-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1794-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1795-2018 | Class II | Terminated | Voluntary: Firm initiated |