510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1539-2018Class II

WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free...

Becton Dickinson & Co / initiated 2018-02-26 / Terminated / root cause: manufacturing process

Beaver-Visitec International Inc. began a recall of WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free, fine dot on the eye that can be used as a precise... on . FDA classified it as Class II and gives the reason as "Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a needle tip instead of the proper flat tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1539-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free, fine dot on the eye that can be used as a precise reference point for a variety of procedures.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a needle tip instead of the proper flat tip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On February 26, 2018, Beaver Visitec International Inc. emailed Product Recall Notices to affected customers. The letter identified the affected device and the reason for the recall. Customers were instructed to destroy affected product, provide evidence of the destruction, and complete a customer response form. Customers will receive a credit. Customers with questions regarding the recall may contact Customer Service at (866)906-8080.
Quantity in commerce
110 total products
Distribution
Distributed to the states of OH, NY, FL, MN, and KS.
Product codes and lots
Lot numbers: 6000292 exp. 7/25/20 and 6000540 exp. 08/30/20  UDI: 30886158014899

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1539-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1539-2018Class IITerminatedVoluntary: Firm initiated