510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2059-2018Class II

RX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical...

Randox Laboratories, Limited / initiated 2018-05-04 / Terminated / root cause: software

Randox Laboratories began a recall of RX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical chemistry tests on . FDA classified it as Class II and gives the reason as "Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2059-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
RX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical chemistry tests.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
All customers will be contacted directly by email and will be followed up with once every 2 weeks over a period of 6 weeks. Customers are advised to update the software in all analyzers. For further questions, please call 1 (866)472-6369.
Quantity in commerce
41 analyzers
Distribution
US Distribution including Puerto Rico and the state of : West Virginia.
Product codes and lots
GTIN 05055273206104. All serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2059-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2059-2018Class IITerminatedVoluntary: Firm initiated