510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1815-2018Class II

Endo GIA" Auto Suture" Universal Articulating Loading Unit 30mm - 2.5mm ( Item Code 030451)

Medtronic, Inc. / initiated 2018-04-26 / Terminated / root cause: manufacturing process

COVIDIEN MEDTRONIC began a recall of Endo GIA" Auto Suture" Universal Articulating Loading Unit 30mm - 2.5mm ( Item Code 030451) Product Usage: The Endo GIA universal and Endo GIA universal XL... on . FDA classified it as Class II and gives the reason as "Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1815-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Endo GIA" Auto Suture" Universal Articulating Loading Unit 30mm - 2.5mm ( Item Code 030451) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the 2.0 mm, 2.5 mm, 3.5 mm or 4.8 mm single use loading unit. The Endo GIA universal staplers will accommodate any of the single use loading unit sizes that are available in the 30 mm, 45 mm and 60 mm lines.
Product code
GDW Staple, Implantable
Reason for recall
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On April 26, 2108, all consignees were issued were issued Urgent Medical Device Recall notices and Return Forms via Federal Express or certified mail. On May 22, 2018, all consignees received an UPDATED URGENT MEDICAL DEVICE RECALL notice. The updated notice included NEW product which has been processed under FDA Ref # 80423. The notice informs customers of the potential for use of a device with a missing sled and the actions they should take to discontinue use of the product and return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. All in-control inventory and devices returned from customers will be reworked. **FOR INTERNATIONAL CUSTOMERS** Please be aware that the recalled units are packed within various convenience kits as identified on the Attachment B provided with your customer letter. Customers are also encouraged to EMAIL OR FAX THE ACKNOWLEDGEMENT FORM TO: - "Product purchased directly from Medtronic": [email protected] or fax to (800) 895-6140. - "Product purchased through distributor": [email protected] or fax it to (203) 492-7719.
Quantity in commerce
171,271 units in total
Distribution
Worldwide and US Nationwide
Product codes and lots
Lots: N7L0071KX, N7L0527KX, N7L0792KX, N7L0914KX, N7L1014KX, N7L1159KX, N8A0154KX,N8A0553KX, N8A0850KX, N8C0830KX, N8C0847KRX, N8C0848KRX  GTIN # 1 (Inner Carton) - 10884523002829 GTIN # 2 (Outer Carton)- 20884523002826

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1815-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1814-2018Class IITerminatedVoluntary: Firm initiated
Z-1815-2018Class IITerminatedVoluntary: Firm initiated
Z-1816-2018Class IITerminatedVoluntary: Firm initiated
Z-1817-2018Class IITerminatedVoluntary: Firm initiated
Z-1818-2018Class IITerminatedVoluntary: Firm initiated
Z-1819-2018Class IITerminatedVoluntary: Firm initiated
Z-1820-2018Class IITerminatedVoluntary: Firm initiated
Z-1821-2018Class IITerminatedVoluntary: Firm initiated
Z-1822-2018Class IITerminatedVoluntary: Firm initiated
Z-1823-2018Class IITerminatedVoluntary: Firm initiated
Z-1824-2018Class IITerminatedVoluntary: Firm initiated
Z-1825-2018Class IITerminatedVoluntary: Firm initiated
Z-1826-2018Class IITerminatedVoluntary: Firm initiated
Z-1828-2018Class IITerminatedVoluntary: Firm initiated
Z-1829-2018Class IITerminatedVoluntary: Firm initiated
Z-1830-2018Class IITerminatedVoluntary: Firm initiated
Z-1831-2018Class IITerminatedVoluntary: Firm initiated
Z-1832-2018Class IITerminatedVoluntary: Firm initiated
Z-1833-2018Class IITerminatedVoluntary: Firm initiated
Z-1834-2018Class IITerminatedVoluntary: Firm initiated
Z-1837-2018Class IITerminatedVoluntary: Firm initiated
Z-1838-2018Class IITerminatedVoluntary: Firm initiated
Z-1839-2018Class IITerminatedVoluntary: Firm initiated
Z-1840-2018Class IITerminatedVoluntary: Firm initiated
Z-1842-2018Class IITerminatedVoluntary: Firm initiated
Z-1843-2018Class IITerminatedVoluntary: Firm initiated
Z-1844-2018Class IITerminatedVoluntary: Firm initiated
Z-1847-2018Class IITerminatedVoluntary: Firm initiated
Z-1848-2018Class IITerminatedVoluntary: Firm initiated
Z-1849-2018Class IITerminatedVoluntary: Firm initiated
Z-1850-2018Class IITerminatedVoluntary: Firm initiated
Z-1852-2018Class IITerminatedVoluntary: Firm initiated
Z-1854-2018Class IITerminatedVoluntary: Firm initiated
Z-1855-2018Class IITerminatedVoluntary: Firm initiated
Z-1858-2018Class IITerminatedVoluntary: Firm initiated
Z-1859-2018Class IITerminatedVoluntary: Firm initiated
Z-1862-2018Class IITerminatedVoluntary: Firm initiated
Z-1863-2018Class IITerminatedVoluntary: Firm initiated
Z-1865-2018Class IITerminatedVoluntary: Firm initiated
Z-1866-2018Class IITerminatedVoluntary: Firm initiated