Z-1597-2018Class II

Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner

Siemens Medical Solutions USA, Inc. / initiated 2018-01-26 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner on . FDA classified it as Class II and gives the reason as "Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1597-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.
Root cause tag
software
FDA root cause
Software Design Change
Action
Customers were notified via letter on approximately 01/26/2018. Instructions included to ensure proper positioning of the PET QC Phantom when performing the PET Quality Control Procedure by following section 6.3.1 in the Biograph Horizon Operator's Manual for Examination and Acquisition, complete the software update to PETsyngo VJ20B, ensure the recall notice is placed in the Biograph Horizon Operator's Manual for Examination and Acquisition and disseminated to all operators of the Biograph Horizon. If the equipment has been sold to another customer, they are instructed to provide the recall notice to the new owner.
Quantity in commerce
42 units
Distribution
Worldwide distribution. US nationwide, Australia, France, India, Japan, and Netherlands.
Product codes and lots
Software version VJ20A. Code No. 10532746, 10532748.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1597-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1597-2018Class IITerminatedVoluntary: Firm initiated