BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*)

bioMerieux, Inc. / initiated 2018-01-04 / Terminated / root cause: design

bioMerieux, Inc. began a recall of BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the... on . FDA classified it as Class II and gives the reason as "Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2429-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV) is an automated test for use on the VIDAS system to detect abnormal operation of the VIDAS and mini VIDAS instrument pipette mechanisms and optical systems. Specifically the mini VIDAS system is an immunodiagnostic system intended to be used by trained and qualified laboratory professionals, for In vitro Diagnostic (IVD) purpose, veterinary and industrial applications. The mini VIDAS system is intended to execute an immunoassay protocol and to release results according to the package insert of the VIDAS assay kits.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS system due to a decrease of pipetted volume.
Root cause tag
design
FDA root cause
Device Design
Action
Subsidiaries and distributors were initially notified of the recall on approximately 01/04/2018. U.S. customers were notified via letter on approximately 01/22/2018. On 03/28/2018 a follow-up notification was issued world-wide to customers to expand the scope. Instructions included to increase the frequency of the Quality Control VIDAS (QCV) testing to weekly rather than monthly to reduce the period of potential retrospective analyses needed, continue to conduct a visual inspection of the SPR after each run (including patient sample test and QCV test) to ensure that the dot on the SPR is still in place, continue to perform External QC testing in accordance with laboratory policy and in compliance with Governmental Regulations & Accreditation Requirements, which includes CLIA, ensure the notification letter is distributed to and reviewed by all appropriate personnel, store the notification letter with the bioMrieux VIDAS and /or miniVIDAS documentation, and complete and return the Acknowledgement Form in Attachment A by Fax to confirm receipt of the notice. For further questions, please call (314) 731-8694
Quantity in commerce
1,866 units
Distribution
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY. International distribution to Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Cape Verde, Chad, Chile, China / Hong Kong, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Djibouti, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Malta, Mauretania, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Tanzania, Thailand, Togo, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zimbabwe.
Product codes and lots
Description, Reference No. (UDI No.): VIDAS ANALYSER 110/220V, 93296 (03573026133573); VIDAS ANALYSER 220V, 93297 (03573026133580); VIDAS ANALYSEUR, 99011 (03573026109257); VIDAS ANALYSER EU, 99444 (03573026114060); VIDAS ANALYSER US, 99445 (03573026114084); VIDAS ANALYSER, 99735 (03573026140427); VIDAS ANALYSER US, W2205 (03573026029678); VIDAS ANALYSER, 410417 (03573026351663); Quality Control VIDAS¿ (QCV), 30706 (03573026121631); VIDAS ANALYSER EU, W3205 (03573026030292); VIDAS INDUSTRY US, 421002 (03573026553319)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2429-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2428-2018Class IITerminatedVoluntary: Firm initiated
Z-2429-2018Class IITerminatedVoluntary: Firm initiated