SOMATOM Force (Model 10742326) Computed tomography x-ray diagnostic system

Siemens Medical Solutions USA, Inc. / initiated 2017-11-08 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force (Model 10742326) Computed tomography x-ray diagnostic system on . FDA classified it as Class II and gives the reason as "There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1748-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Force (Model 10742326) Computed tomography x-ray diagnostic system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM Definition Edge, SOMATOM Definition Flash and SOMATOM Force.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens Medical Solutions USA, Inc. will bring defect into Compliance noted below: 1.You will contact customers and initiate the actions to correct the failure to comply issue. 2. You will provide a software update to correct tube current calculations by the CARE Dose4D algorithm for head scans. 3.You will follow-up with consignees to verify the effectiveness of the software update. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) division office coordinator noted below are to be included. For further questions, please call (610) 448-6471.
Quantity in commerce
91
Distribution
Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico.
Product codes and lots
Serial Numbers: 75708 75765 75714 75790 75709 75607 75784 75789 75617 75550 75526 75512 75775 75567 75536 75807 75711 75716 75514 75770 75600 75794 75563 75824 75768 75510 75792 75454 75754 75755 75540 75739 75806 75808 75571 75818 75437 75500 75527 75776 75728 75475 75535 75820 75727 75481 75439 75450 75482 75760 75710 75460 75476 75588 75493 75458 75528 75467 75823 75761 75602 75487 75804 75837 75743 75787 75793 75478 75604 75805 75513 75532 75843 75555 75788 75524 75778 75515 75559 75610 75501 75570 75769 75774 75752 75799 75616 75750 75576 75845 75605

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1748-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1745-2018Class IITerminatedVoluntary: Firm initiated
Z-1746-2018Class IITerminatedVoluntary: Firm initiated
Z-1747-2018Class IITerminatedVoluntary: Firm initiated
Z-1748-2018Class IITerminatedVoluntary: Firm initiated