Z-2100-2018Class II

VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit

VGI medical, LLC / initiated 2018-04-09 / Terminated / root cause: manufacturing process

VGI Medical, LLC began a recall of VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit on . FDA classified it as Class II and gives the reason as "IFU contained an automated cleaning process for the VerteLP Instrument Tray that was not validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2100-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
VGI medical, LLC
Product
VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit
Product code
OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar
Reason for recall
IFU contained an automated cleaning process for the VerteLP Instrument Tray that was not validated
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm notified their direct consignees by letter on 04/09/2018 and followed with an e-mail on 04/11/2018. The letter directed the direct consignees (distributors) to instruct their accounts (user) to discontinue the automated cleaning process and to provide users with the new IFUs.
Quantity in commerce
6 units
Distribution
Distributed to MI, FL, and NY.
Product codes and lots
Lots of Instrument Sets: LP-INSTSET-001, VLP-INSTSET-002, VLP-INSTSET-003, VLP-INSTSET-004, VLP-INSTSET-005, VLP-INSTSET-006.  Lots of individual instruments contained in sets: 127617, 127901, 127902, 127903, 127904, 127905, 127906, 127907, 127908, 133003, 133004, 133005, 1330-1, 1330-2, 1330-6, 133004, 1285001, 144301, 144302, 163214, 1401001, 157801, 162402 163203, 163204, 163205, 163206, 163207, 163208, 163209, 163212, 163213, and 163214.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2100-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2100-2018Class IITerminatedVoluntary: Firm initiated