Z-2100-2018Class II
VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit
VGI medical, LLC / initiated 2018-04-09 / Terminated / root cause: manufacturing process
VGI Medical, LLC began a recall of VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit on . FDA classified it as Class II and gives the reason as "IFU contained an automated cleaning process for the VerteLP Instrument Tray that was not validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2100-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- VGI medical, LLC
- Product
- VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit
- Product code
- OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar
- Reason for recall
- IFU contained an automated cleaning process for the VerteLP Instrument Tray that was not validated
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The firm notified their direct consignees by letter on 04/09/2018 and followed with an e-mail on 04/11/2018. The letter directed the direct consignees (distributors) to instruct their accounts (user) to discontinue the automated cleaning process and to provide users with the new IFUs.
- Quantity in commerce
- 6 units
- Distribution
- Distributed to MI, FL, and NY.
- Product codes and lots
- Lots of Instrument Sets: LP-INSTSET-001, VLP-INSTSET-002, VLP-INSTSET-003, VLP-INSTSET-004, VLP-INSTSET-005, VLP-INSTSET-006. Lots of individual instruments contained in sets: 127617, 127901, 127902, 127903, 127904, 127905, 127906, 127907, 127908, 133003, 133004, 133005, 1330-1, 1330-2, 1330-6, 133004, 1285001, 144301, 144302, 163214, 1401001, 157801, 162402 163203, 163204, 163205, 163206, 163207, 163208, 163209, 163212, 163213, and 163214.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2100-2018 | Class II | Terminated | Voluntary: Firm initiated |