Z-1954-2018Class II

Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a...

Draeger, Inc. / initiated 2018-04-26 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel on . FDA classified it as Class II and gives the reason as "The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1954-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On April 27, 2018, the recalling firm contacted its consignees by telephone prior to sending out Urgent Medical Device Recall notifications via e-mail. Customers were instructed to remove affected devices from operation immediately. The firm provided replacement devices so customers could send their affected ventilator to the firm's repair center for update once the required parts were available. Customers were also asked to inform all affected users at each site. If you have any questions regarding this letter please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437. For questions regarding the operation and/or servicing of your Carina Ventilator please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437.
Quantity in commerce
8 units
Distribution
Distributed to the states of CT, IL, MN, NV, SD, and WA.
Product codes and lots
Serial nos. ASKM-0065 ASKM-0066 ASKM-0106 ASKM-0134 ASKM-0137 ASKM-0150 ASKM-0151 ASKN-0009

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1954-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1954-2018Class IITerminatedVoluntary: Firm initiated