Z-1903-2018Class I

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller...

HeartWare, Inc. / initiated 2018-05-02 / Open, Classified / root cause: design

Heartware began a recall of Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420... on . FDA classified it as Class I and gives the reason as "Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1903-2018
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartWare, Inc.
Product
Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.
Root cause tag
design
FDA root cause
Device Design
Action
The firm notified physicians and healthcare professionals about the device correction by letter on 05/02/2018. The letter explained problem and provided recommendations for effective power source management of the system. Medtronic began notifying VAD Clinicians in the US on May 31, 2018, mainly via 2-day courier, and OUS via locally approved methods, of a newly available field service. A field service for application of a lubricant to the HVAD System power source connectors to mitigate the potential for transient interruptions of the electrical connection between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the HVAD Controller is now available.
Quantity in commerce
175878 devices
Distribution
Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, and Vietnam.
Product codes and lots
ALL SERIAL NUMBERS  UPN/GTIN: (a) 1400 and 1401: distributed prior to GTN requirement  1403: 00888707000116, 00888707000475  1407: '00888707000291, '00888707000727, '00888707000710,   00888707000734, '00888707000741, '00888707000765, '00888707000758, '00888707000789, '00888707000772, '00888707000499, '00888707000796, '00888707000802, '00888707000819, '00888707000482, '00888707004534, '0888707001670, '00888707001656, '00888707001663, '00888707001670, '00888707001687, '00888707001694, '00888707001700, '00888707001717, '00888707001724, '00888707001731, '00888707001748, '00888707001755, '00888707001762, '00888707002813, '00888707002820, '00888707002837, '00888707002844, '00888707002851, '00888707001472  1420: '00888707000420, '00888707000437, '00888707002660  (b) 1435: '00888707000109  1440: '00888707000260, '00888707001885, '00888707001496,  '00888707002745  (c) 1425: '00888707000093  1430: '00888707000307, '00888707000826, '00888707000833,  '00888707000840, '00888707000857, '00888707000871, '00888707000864,  '00888707000888, '00888707000901, '00888707000505, '00888707000895,  '00888707000918, '00888707001489, '00888707001779, '00888707001786,  '00888707001793, '00888707001809, '00888707001816, '00888707001823,  '00888707001830, '00888707001847, '00888707001854, '00888707001861,  '00888707001878, '00888707002738, '00888707002769, '00888707002776,  '00888707002783, '00888707002790, '00888707002806, '00888707004817  (d) 1650DE: 00888707000369, 00888707001373, '00888707001588, '00888707001366, '00888707002684 '00888707002646, '00888707002653,  '00888707000376

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1903-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1903-2018Class IOngoingVoluntary: Firm initiated