Z-2143-2018Class II

IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865352

Philips North America LLC / initiated 2018-04-30 / Terminated / root cause: labeling and UDI

Philips Electronics North America Corporation began a recall of IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865352 on . FDA classified it as Class II and gives the reason as "Five warning statements are missing from the instructions for use". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2143-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865352
Product code
DSI Detector And Alarm, Arrhythmia
Reason for recall
Five warning statements are missing from the instructions for use.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Urgent Medical Device Correction notification letters were distributed on 4/30/18. The notification included the following instructions: Customers were asked to please review the information provided in the letter with all members of their staff who need to be aware of its contents. The letter provides information on the missing warnings and instructions for actions to be taken. The enclosed MX40 IFU Errata Sheet must be attached to the first page of Chapter 6 of the Instructions for Use for ready reference. Complete and return the attached Customer Reply Form.
Quantity in commerce
4,488
Distribution
Worldwide Distribution -- US, to the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and to the territories of Guam and Puerto Rico; and, countries of Argentina, Australia, Austria, Belgium, Canada, Chile, China, Croatia, Czech Republic, Denmark, Estonia, Finland, France, German, Ghana, Greece, Guadeloupe, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Malaysia, Martinique, Mexico, Myanmar, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UK, UAE, and Vietnam.
Product codes and lots
All devices since release of B.05 (March 2016) prefix date code >/= US096

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2143-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2141-2018Class IITerminatedVoluntary: Firm initiated
Z-2142-2018Class IITerminatedVoluntary: Firm initiated
Z-2143-2018Class IITerminatedVoluntary: Firm initiated
Z-2144-2018Class IITerminatedVoluntary: Firm initiated