Z-1930-2018Class II

Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of...

Siemens Medical Solutions USA, Inc. / initiated 2018-02-21 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of high resolution physiologic and anatomic information on . FDA classified it as Class II and gives the reason as "Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a waveform sampling issue caused by a firmware change within the Universal...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1930-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a waveform sampling issue caused by a firmware change within the Universal Physiological Measurement Module.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Customers were notified via letter on approximately 02/21/2018. Letters were sent via certified mail or email. Instructions included to discontinue use of the system (according to the cardiac imaging protocols listed in the letter), ensure the customer advisory notice is placed in the Biograph Horizon Operator's Manual, ensure the recall information is disseminated to all operators of the system, coordinate the installation of the updated Universal Physiological Measurement Module with Siemens Healthineers, and if the equipment has been sold to forward the letter to the new owner.
Quantity in commerce
14 units
Distribution
Domestic distribution to . International distribution to Finland, France, France, Iran, Ireland, Italy, Japan, Japan, Netherlands.
Product codes and lots
Code No. 10532746, 10532748

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1930-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1930-2018Class IITerminatedVoluntary: Firm initiated