Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz...
LivaNova USA began a recall of Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz on . FDA classified it as Class II and gives the reason as "Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1901-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 04/20/2018, a Medical Device Correction notice was sent to customers informing them to return devices less than ten years old. Returned devices will receive a deep cleaning service, which will reduce microbial contamination to HPC < 100 CFU/ml of water and NTM < 1 CFU/100ml.
- Quantity in commerce
- 2408
- Distribution
- MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,
- Product codes and lots
- Serial #: a) 16S10743 - 6S11708 b) 16S10772 - 16S16840 C) 16S10958 - 16S16847
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1901-2018 | Class II | Terminated | Voluntary: Firm initiated |