510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1901-2018Class II

Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz...

LivaNova USA, Inc. / initiated 2018-04-20 / Terminated / root cause: design

LivaNova USA began a recall of Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz on . FDA classified it as Class II and gives the reason as "Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1901-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
Root cause tag
design
FDA root cause
Device Design
Action
On 04/20/2018, a Medical Device Correction notice was sent to customers informing them to return devices less than ten years old. Returned devices will receive a deep cleaning service, which will reduce microbial contamination to HPC < 100 CFU/ml of water and NTM < 1 CFU/100ml.
Quantity in commerce
2408
Distribution
MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,
Product codes and lots
Serial #: a) 16S10743 - 6S11708  b) 16S10772 - 16S16840  C) 16S10958 - 16S16847

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1901-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1901-2018Class IITerminatedVoluntary: Firm initiated