Z-2175-2018Class II

Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray...

Philips North America LLC / initiated 2018-03-30 / Terminated / root cause: design

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2175-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Field Safety Notices dated 3/28/18 were distributed to customers. The letters instructed customers to perform the following actions: Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. As outlined in the Instructions for Use, always monitor the patient during all movements (manual or motorized) of the patient support. If you recognize an impending collision with the back shelter wall, halt the motion or activate the system E-stop. Mandatory Field Change Order (FCO) 72800686 which will add additional warning labels specific to this issue to the system. The FCO will be implemented free of charge.
Quantity in commerce
2
Distribution
Worldwide distribution including US states of CA, MD, NV, OH, TX, and WI, Afghanistan, Canada, Germany, Iraq, Italy, Korea, Kuwait, and Lebanon.
Product codes and lots
System Serial Numbers: 6147, 6143

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2175-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2174-2018Class IITerminatedVoluntary: Firm initiated
Z-2175-2018Class IITerminatedVoluntary: Firm initiated
Z-2176-2018Class IITerminatedVoluntary: Firm initiated