MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease Handpiece...
MicroAire Surgical Instruments, LLC began a recall of MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease Handpiece, containing IFU rev. D The MicroAire Carpal Tunnel Release... on . FDA classified it as Class II and gives the reason as "The Instructions for Use (IFU) is incorrect with regards to sterilization parameters and must be replaced". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2107-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MicroAire Surgical Instruments, LLC
- Product
- MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease Handpiece, containing IFU rev. D The MicroAire Carpal Tunnel Release System is for use in patients diagnosed with carpal tunnel syndrome that is not associated with, or secondary to, any other known pathology (i.e., idiopathic carpal tunnel syndrome). Preoperative x-rays of the wrist, including a carpal tunnel view, are recommended to aid in the diagnosis of associated pathology (i.e., calcific tendonitis, fracture of the hook of the hamate). This device is indicated for releasing the transverse carpal ligament (flexor retinaculum).
- Product code
- HRX Arthroscope
- Reason for recall
- The Instructions for Use (IFU) is incorrect with regards to sterilization parameters and must be replaced.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Urgent Medical Device Recall letters were distributed to customers. The letters instructed customers to do the following: If you have some quantity of IFUs remaining, we're asking you to destroy your current copies, IM-SMARTRELEASE Rev. A-D, and replace them by going on-line at MicroAire at www.microaire.com/resources. Click Carpal Tunnel, then SmartRelease Instruction Manual to obtain the latest version, IM-SMARTRELEASE Rev. E. Please complete and send back page two of this letter to [email protected] or via fax at 800-648-4309. If you have no quantity of IFUs remaining, there are no further activities needed by you except for the mandatory returning of page two of this letter to [email protected] or via fax at 800-648-4309. For further questions, please call (800) 722-0822.
- Quantity in commerce
- 2009
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of: Australia, Belgium, Brazil, Canada, Chile, Czech Republic, Denmark, France, Germany, Hong Kong, India, Iran, Ireland, Israel, Italy, Jordan, Korea, Lebanon, Mexico, Nicaragua, Poland, Russia, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, and UK.
- Product codes and lots
- All products distributed from 1/1/13 to 2/23/18
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2107-2018 | Class II | Terminated | Voluntary: Firm initiated |