Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
TriMed Inc. began a recall of Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19 on . FDA classified it as Class II and gives the reason as "During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2262-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Trimed Inc
- Product
- Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
- Product code
- JDR Staple, Fixation, Bone
- Reason for recall
- During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.
- Quantity in commerce
- 532
- Distribution
- U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
- Product codes and lots
- Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2262-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2263-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2264-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2265-2018 | Class II | Terminated | Voluntary: Firm initiated |