Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19

Trimed Inc / initiated 2018-04-13 / Terminated / root cause: manufacturing process

TriMed Inc. began a recall of Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19 on . FDA classified it as Class II and gives the reason as "During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2262-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Trimed Inc
Product
Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
Product code
JDR Staple, Fixation, Bone
Reason for recall
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.
Quantity in commerce
532
Distribution
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Product codes and lots
Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ,  10505-23, 050523, 180502, 180901,  180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805,  180502,  118901

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2262-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2262-2018Class IITerminatedVoluntary: Firm initiated
Z-2263-2018Class IITerminatedVoluntary: Firm initiated
Z-2264-2018Class IITerminatedVoluntary: Firm initiated
Z-2265-2018Class IITerminatedVoluntary: Firm initiated