Z-2110-2018Class II
NaturaLyte Liquid Acid Concentrate for Bicarbonate Dialysis, 3.43 Liters, Part Number: 08-3231-1
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2018-05-02 / Terminated / root cause: manufacturing process
Fresenius Medical Care Renal Therapies Group, LLC began a recall of NaturaLyte Liquid Acid Concentrate for Bicarbonate Dialysis, 3.43 Liters, Part Number: 08-3231-1 Product Usage: This concentrate is formulated to be used in... on . FDA classified it as Class II and gives the reason as "The bottles are mislabeled with an incorrect part number". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2110-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- NaturaLyte Liquid Acid Concentrate for Bicarbonate Dialysis, 3.43 Liters, Part Number: 08-3231-1 Product Usage: This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine. For every 45 volume parts of dialysate: 1 volume part of the NaturaLyte acid concentrate, 1.72 volume parts of FMCNA Sodium Bicarbonate Concentrate and 42.28 volume parts of Purified Water (AAMI quality) are used.
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- The bottles are mislabeled with an incorrect part number.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Phone calls and Urgent Medical Device Recall notification letters were issued on 5/2/18. Customers were instructed to do the following: Immediately examine your stock to determine whether you have any NaturaLyte Liquid Acid from the lot listed above. If any products of this lot is found, discontinue use immediately. Place all units in a secure, segregated area. Contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product and arrange for replacement product. Promptly fill out and return the attached reply form.
- Quantity in commerce
- 1060 cases (4240 bottles)
- Distribution
- Worldwide Distribution - US Nationwide in the states: CA, IA, IL, IN, KS, MD, MI, MN, MS, NE, NJ, NV, OK, PA, SD, TX, and WI.
- Product codes and lots
- Lot Code: 18CXAC095
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2110-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY