Z-2168-2018Class II

The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical...

bioMerieux, Inc. / initiated 2018-03-23 / Terminated / root cause: manufacturing process

BioMerieux SA began a recall of The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the... on . FDA classified it as Class II and gives the reason as "A shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2168-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram negative bacilli to antimicrobial agents. Fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
A shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
The firm, Biomerieux, sent an "Urgent product removal notice" letter on April 23 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Immediately acknowledge receipt (AR) of this FSCA. 2. Identify all countries and customers for which you are responsible that are impacted by this FSCA. 3. Define additional actions required (e.g. Translate customer letter, send to impacted countries/customers, etc.) 4. Distribute the customer letter to all customers that have received the products 5. Discard any impacted product remaining in local inventory 6. After all actions above are complete, please return the acknowledgement of completion (AC) for this FSCA. The due date for completion of the required actions and submitting the AC is 23-APR-2018 Subsidiaries Subsidiaries using SAP, please manage both the Acknowledgment of Receipt and the Acknowledgment of Completion of activities in SAP. Subsidiaries not using SAP, please immediately upload the Acknowledgement of Receipt in LiveLink after signing and completing sections #1 and #2. Section #3 must be signed, completed and uploaded in LiveLink before the due date. Distributors Please immediately Acknowledge Receipt of this FSCA by completing sections #1 and #2 of the acknowledgement form and sending it by email to [email protected] or by fax to +33 4 78 87 21 79. Please remember to sign and complete all of sections #1 and #2. Sections #1 and #3 must be signed, completed, faxed or emailed before the due date. If you have any questions, contact the Specialist, Regulatory & Quality Compliance at +33 4 78 87 56 16 or email to: [email protected].
Quantity in commerce
135 (78 boxes of various product) in total
Distribution
International distribution.
Product codes and lots
VITEK¿ 2 AST-N233 (ref: 413117): Lot Number 6330343103

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2168-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2164-2018Class IITerminatedVoluntary: Firm initiated
Z-2165-2018Class IITerminatedVoluntary: Firm initiated
Z-2166-2018Class IITerminatedVoluntary: Firm initiated
Z-2167-2018Class IITerminatedVoluntary: Firm initiated
Z-2168-2018Class IITerminatedVoluntary: Firm initiated
Z-2169-2018Class IITerminatedVoluntary: Firm initiated
Z-2170-2018Class IITerminatedVoluntary: Firm initiated
Z-2171-2018Class IITerminatedVoluntary: Firm initiated
Z-2172-2018Class IITerminatedVoluntary: Firm initiated