The PREVI Color Gram dyes are used with the PREVI Color Gram instrument to stain bacterial and fungal...
BioMerieux SA began a recall of The PREVI Color Gram dyes are used with the PREVI Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram... on . FDA classified it as Class II and gives the reason as "A shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2170-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- The PREVI Color Gram dyes are used with the PREVI Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI Color Gram instrument sprays the PREVI Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.
- Product code
- KPA Slide Stainer, Automated
- Reason for recall
- A shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- The firm, Biomerieux, sent an "Urgent product removal notice" letter on April 23 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Immediately acknowledge receipt (AR) of this FSCA. 2. Identify all countries and customers for which you are responsible that are impacted by this FSCA. 3. Define additional actions required (e.g. Translate customer letter, send to impacted countries/customers, etc.) 4. Distribute the customer letter to all customers that have received the products 5. Discard any impacted product remaining in local inventory 6. After all actions above are complete, please return the acknowledgement of completion (AC) for this FSCA. The due date for completion of the required actions and submitting the AC is 23-APR-2018 Subsidiaries Subsidiaries using SAP, please manage both the Acknowledgment of Receipt and the Acknowledgment of Completion of activities in SAP. Subsidiaries not using SAP, please immediately upload the Acknowledgement of Receipt in LiveLink after signing and completing sections #1 and #2. Section #3 must be signed, completed and uploaded in LiveLink before the due date. Distributors Please immediately Acknowledge Receipt of this FSCA by completing sections #1 and #2 of the acknowledgement form and sending it by email to [email protected] or by fax to +33 4 78 87 21 79. Please remember to sign and complete all of sections #1 and #2. Sections #1 and #3 must be signed, completed, faxed or emailed before the due date. If you have any questions, contact the Specialist, Regulatory & Quality Compliance at +33 4 78 87 56 16 or email to: [email protected].
- Quantity in commerce
- FUCHSIN-A (ref: 29522) 15 eaches; IODINE-B (ref: 29523) 10 eaches ; CRYSTAL VIOLET-C (ref: 29524) 10 eaches
- Distribution
- International distribution.
- Product codes and lots
- Reagents for use with the PREVI¿ Color Gram: IODINE-B (ref: 29523) Lot number: G59407 Exp. 15-Aug-18 CRYSTAL VIOLET-C (ref: 29524) Lot number: G52306 Exp.19-Jul-18 FUCHSIN-A (ref: 29522): Lot Number Lot number: F89544 Exp. 13-Apr-18
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2164-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2165-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2166-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2167-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2168-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2169-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2170-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2171-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2172-2018 | Class II | Terminated | Voluntary: Firm initiated |