Femoral Nail, A/R T2 Femur 10x320 mm, Product Number 18251032S
Stryker GmbH began a recall of Femoral Nail, A/R T2 Femur 10x320 mm, Product Number 18251032S on . FDA classified it as Class II and gives the reason as "The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2394-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Femoral Nail, A/R T2 Femur 10x320 mm, Product Number 18251032S
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Urgent Product Recall letters dated 4/27/18 were distributed to customers. The letters instructed customers to perform the following actions: 1. Inform individuals within your organization who need to be aware of this device recall. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. Response is required, even though you may not have any physical inventory on site anymore. 3. Quarantine and discontinue use of the recalled devices. 4. Maintain awareness of this notice internally until all required actions have been completed within your facility 5. Inform Stryker if any of the subject devices have been distributed to other organizations. a) Please provide contact details so that Stryker can inform the recipients appropriately. b) If you are a Distributor, note that you are responsible for notifying your affected customers. 6. Branches/Agencies: Return all affected products available at your location to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 PR 1774317 Event 6020 7. Hospitals/Branches/Agencies: Complete and sign the enclosed Business Reply Form and fax to 1-888-230-2125 or to Email: [email protected]
- Quantity in commerce
- 2
- Distribution
- The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.
- Product codes and lots
- GTIN 04546540199492; Lot code: K081642
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2393-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2394-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2395-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2396-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2397-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2398-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2399-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2400-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2401-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2402-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2403-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2404-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2405-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2406-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2407-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2408-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.