Space Pump IV Set, 2 Caresites, ASV, 0.2 Micron Air Eliminating Filter, Product Code 363420 The Infusomat...
B. Braun Medical, Inc. began a recall of Space Pump IV Set, 2 Caresites, ASV, 0.2 Micron Air Eliminating Filter, Product Code 363420 The Infusomat Space Pump IV Set, 2 Caresites, ASV with 0.2 Micron... on . FDA classified it as Class II and gives the reason as "There is a potential for a different set to be mixed within the lot, where the incorrect set does not contain an air eliminating filter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2285-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Space Pump IV Set, 2 Caresites, ASV, 0.2 Micron Air Eliminating Filter, Product Code 363420 The Infusomat Space Pump IV Set, 2 Caresites, ASV with 0.2 Micron Air Eliminating Filter and Anti-Syphon Valve (ASV) is a latex-free product intended to be used with the Infusomat Space Pump, to deliver intravenous solutions. The 0.2 Micron filter is a clear acrylic in-line filter designed to eliminate air within the IV infusion set. While it is not listed in the labeling, it is known that the filter is used for removal of particulate matter for some medications and at risk patient populations.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- There is a potential for a different set to be mixed within the lot, where the incorrect set does not contain an air eliminating filter.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- B. Braun Medical Inc. sent an Urgent Voluntary Medical Device Recall Notification letter dated May 9, 2018. The letter included the following instructions for customers: 1. Review the Medical Device Recall Notification in its entirety and ensure that all users in your organization of the above mentioned product, and other concerned persons, are informed about this voluntary product recall and the affected product. Further use of this product should be discontinued immediately and affected product quarantined. 2. If you are a distributor, please immediately forward this recall notification to your customers. 3. Determine your current inventory of the affected lots within your facility. Do not destroy any affected product. 4. Utilizing the attached "Product Removal Acknowledgement" form, record the total number of individual units (within partial cases) and the number of full-unopened cases. If you have no inventory remaining, please enter zero (0) on the form. 5. Return the completed "Product Removal Acknowledgement" form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). It is important this form is returned, so BBMI can meet its United States Food and Drug Administration regulatory requirement. 6. If you have any full cases, partial cases or unused individual pieces of these affected products as identified in the " Product Removal Acknowledgement" form that was submitted to BBMI Quality Assurance Department, a BBMI Customer Support Representative will contact you to provide instructions for handling the affected product and arrange for return to BBMI. This may include credit or replacement product. For further questions, please call (610) 596-0500.
- Quantity in commerce
- 1,872
- Distribution
- US Distribution to the states of : CA, IL, and TX., and Internationally to Canada.
- Product codes and lots
- Lot Number: 0061583973
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2285-2018 | Class II | Terminated | Voluntary: Firm initiated |