Z-2344-2018Class II

VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic

Ortho-Clinical Diagnostics, Inc. / initiated 2018-05-09 / Terminated

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2344-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic
Product code
DIP Enzyme Immunoassay, Diphenylhydantoin
Reason for recall
There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.
FDA root cause
Under Investigation by firm
Action
Urgent Product Correction Notification letters were distributed on 5/9/18. The letters instruct customers to do the following: " Immediately discontinue using and discard all remaining inventory of VITROS PHYT Slides for the affected lots. " Complete the Confirmation of Receipt form no later than May 23, 2018. Ortho will replace your remaining inventory of the affected lots or credit your account. " Post this notification by each VITROS System in your facility or with the user documentation. " Forward this notification if the product was distributed outside of your facility. If the customer is a distributor, they were also instructed to: " Send the letter to all customers who were shipped an affected lot of VITROS PHYT Slides.
Quantity in commerce
5712
Distribution
Worldwide distribution. US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, and Sweden
Product codes and lots
Lot Numbers: 2616-0165-4866, 2616-0165-4904, 2616-0165-4870, 2616-0165-4869, 2616-0165-5266, 2616-0165-5307, 2616-0165-5233, 2616-0165-5232, 2616-0165-5611, 2616-0165-5610, 2616-0165-5648, 2616-0165-5649, 2616-0165-6218, 2616-0165-6275, 2616-0165-6220, 2616-0165-6221, 2616-0165-6219, 2616-0165-6276, 2616-0165-6791, 2616-0165-6832, 2616-0165-6792, 2616-0165-6787, 2616-0165-7225, 2616-0165-7226, 2616-0165-7223, 2616-0165-7214, 2616-0165-7222

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2344-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2344-2018Class IITerminatedVoluntary: Firm initiated