VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic
Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2344-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic
- Product code
- DIP Enzyme Immunoassay, Diphenylhydantoin
- Reason for recall
- There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.
- FDA root cause
- Under Investigation by firm
- Action
- Urgent Product Correction Notification letters were distributed on 5/9/18. The letters instruct customers to do the following: " Immediately discontinue using and discard all remaining inventory of VITROS PHYT Slides for the affected lots. " Complete the Confirmation of Receipt form no later than May 23, 2018. Ortho will replace your remaining inventory of the affected lots or credit your account. " Post this notification by each VITROS System in your facility or with the user documentation. " Forward this notification if the product was distributed outside of your facility. If the customer is a distributor, they were also instructed to: " Send the letter to all customers who were shipped an affected lot of VITROS PHYT Slides.
- Quantity in commerce
- 5712
- Distribution
- Worldwide distribution. US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, and Sweden
- Product codes and lots
- Lot Numbers: 2616-0165-4866, 2616-0165-4904, 2616-0165-4870, 2616-0165-4869, 2616-0165-5266, 2616-0165-5307, 2616-0165-5233, 2616-0165-5232, 2616-0165-5611, 2616-0165-5610, 2616-0165-5648, 2616-0165-5649, 2616-0165-6218, 2616-0165-6275, 2616-0165-6220, 2616-0165-6221, 2616-0165-6219, 2616-0165-6276, 2616-0165-6791, 2616-0165-6832, 2616-0165-6792, 2616-0165-6787, 2616-0165-7225, 2616-0165-7226, 2616-0165-7223, 2616-0165-7214, 2616-0165-7222
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2344-2018 | Class II | Terminated | Voluntary: Firm initiated |