510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2183-2018Class II

Vanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement...

Zimmer GmbH / initiated 2018-05-11 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of Vanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement procedures on . FDA classified it as Class II and gives the reason as "The label master file was errantly set up as a 63/37mm instead of a 63/67mm. There is no 63/37 size offered, and the product is laser marked with the correct size". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2183-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Vanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement procedures.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The label master file was errantly set up as a 63/37mm instead of a 63/67mm. There is no 63/37 size offered, and the product is laser marked with the correct size.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
The firm, Zimmer Biomet, sent an "Urgent Medical Device Recall (Removal)" letters dated 5/11/2018 to customers on 5/11/18. The letters described the product, problem and actions to be taken. The letters instruct customers to perform the following actions: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form. a. Return a digital copy to [email protected] within three (3) days. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form to Zimmer Biomet. a. For each return, send a copy of Attachment 1 to [email protected]. b. Include a hardcopy of Attachment 1 with your shipment for immediate processing. c. Mark the outside of the returning product box(es) clearly with RECALL. 5. Distributors will notify their accounts to which the product was further distributed and to locate and return products from these accounts in accordance with local regulations. 6. Retain a copy of your field action acknowledgement and product return forms for your records in the event of a compliance audit of your facility. 7. If after reviewing this notice you have further questions or concerns please call the customer call center at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voice mail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
169
Distribution
Worldwide distribution: US (nationwide) to states of: AR, CA, FL, GA, IL, IN, KS, MI, MN, MS, NC, OH, PA, TN, TX, and WV and country of: Canada.
Product codes and lots
UDI: (01)00880304673885(17)230222(10)120150, Lot Number: 120150;  UDI: (01)00880304673885(17)201203(10)140700, Lot Number: 140700; UDI: (01)00880304673885(17)211125(10)173480, Lot Number: 173480;  UDI: (01)00880304673885(17)210520(10)205730, Lot Number: 205730;  UDI: (01)00880304673885(17)210822(10)294370, Lot Number: 294370;  UDI: (01)00880304673885(17)210310(10)338750, Lot Number: 338750;  UDI: (01)00880304673885(17)210902(10)421970, Lot Number: 421970;  UDI: (01)00880304673885(17)210916(10)516680, Lot Number: 516680;  UDI: (01)00880304673885(17)210411(10)678370, Lot Number: 678370;  UDI: (01)00880304673885(17)220603(10)700470, Lot Number: 700470;  UDI: (01)00880304673885(17)201015(10)722060, Lot Number: 722060;  UDI: (01)00880304673885(17)230130(10)775330, Lot Number: 775330;  UDI: (01)00880304673885(17)220620(10)858220, Lot Number: 858220;  UDI: (01)00880304673885(17)211101(10)888080, Lot Number: 888080;  UDI: (01)00880304673885(17)220221(10)992290, Lot Number: 992290

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2183-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2183-2018Class IITerminatedVoluntary: Firm initiated