Philips Ingenuity Core 728321
Philips Medical Systems (Cleveland) Inc began a recall of Philips Ingenuity Core 728321 on . FDA classified it as Class II and gives the reason as "A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2274-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Ingenuity Core 728321
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Urgent Field Safety Notice letters were sent to customers on 3/28/18. The letters did not request any actions to be conducted by the customers. For Brilliance Big Bore/ Ingenuity CT / Ingenuity Core / Ingenuity Corel 28 Customers: A Philips Field Service Engineer will contact you to schedule the FCO updates at your site. Field Change Order (FCO 72800691 and 72800692) will be released to correct the issue. For Ingenuity TF PET/CT Customers: Field Change Order (FCO 88200520 and 88200511) will be released to correct this issue. A Philips Field Service Engineer will contact you to schedule the FCO updates at your site.
- Quantity in commerce
- 28 total
- Distribution
- The systems were distributed to the following US states: HI, KS, and MO. The systems were distributed to the following foreign countries: Algeria, Austria, Brazil, China, Indonesia, Japan, Kenya, Korea, Mexico, Poland, Turkey, and Uganda.
- Product codes and lots
- UDI: (01)00884838059498(21)333169, Serial Number: 333169; UDI: (01)00884838059498(21)310377, Serial Number: 310377; UDI: (01)00884838059498(21)310375, Serial Number: 310375; UDI: (01)00884838059498(21)333171, Serial Number: 333171; UDI: (01)00884838059498(21)333172, Serial Number: 333172
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2273-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2274-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2275-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2276-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2277-2018 | Class II | Terminated | Voluntary: Firm initiated |