510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2239-2018Class II

BEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated...

Zimmer GmbH / initiated 2018-05-16 / Terminated / root cause: packaging

Zimmer Biomet, Inc. began a recall of BEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail... on . FDA classified it as Class II and gives the reason as "The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2239-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
BEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Product code
HTY Pin, Fixation, Smooth
Reason for recall
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 16, 2018 on May 16, 2018, to affected consignees via email and FedEx. The letter describes the product, problem and actions to be taken. Consignees were informed of the recall and instructed to take the following actions: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. For each return, send a copy of Attachment 1 to [email protected] b. International returns, please request an IRA by emailing [email protected] c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals included with the email notification sent to your facility, which includes a list of hospitals that have already been notified of this recall. b. Identify whether there are any additional hospitals that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format
Quantity in commerce
2168
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.
Product codes and lots
563080 563090 407870 408160 408140 412690 412750 412720 412740 668840 202750 389030 410170 410180 395170 513920 201790 232930 639560 800410 980930 134560 271820 351760 504870 578400 605700 750960 750970 913730 921150 058010 631570 969080 101380 944200 969090 203000 449730 449740 449750 449720 665510 820130 850170 978340 221240 419010 848820 848810 278720 387620 437340 502200 623450 038560 186020 212630 183170 183180 183190 434630 183200 930950 693480

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2239-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2230-2018Class IITerminatedVoluntary: Firm initiated
Z-2231-2018Class IITerminatedVoluntary: Firm initiated
Z-2232-2018Class IITerminatedVoluntary: Firm initiated
Z-2233-2018Class IITerminatedVoluntary: Firm initiated
Z-2234-2018Class IITerminatedVoluntary: Firm initiated
Z-2235-2018Class IITerminatedVoluntary: Firm initiated
Z-2236-2018Class IITerminatedVoluntary: Firm initiated
Z-2237-2018Class IITerminatedVoluntary: Firm initiated
Z-2238-2018Class IITerminatedVoluntary: Firm initiated
Z-2239-2018Class IITerminatedVoluntary: Firm initiated
Z-2240-2018Class IITerminatedVoluntary: Firm initiated
Z-2241-2018Class IITerminatedVoluntary: Firm initiated
Z-2242-2018Class IITerminatedVoluntary: Firm initiated