Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2339-2018Class II

Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile

Trilliant Surgical LLC / initiated 2018-03-27 / Terminated

Trilliant Surgical Ltd. began a recall of Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile on . FDA classified it as Class II and gives the reason as "Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2339-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Trilliant Surgical LLC
Product
Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile
Product code
MDM Instrument, Manual, Surgical, General Use
Reason for recall
Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.
FDA root cause
Other
Action
On 03/27/18, the firm sent Medical Device Recall notices to customers informing them to: 1) Check inventory stock for identified devices; 2) Locate and remove and/or discontinue use of identified devices; 3) Return all related product(s) to the firm in accordance with an included return response / acknowledgement form. Customers were instructed to contact the firm at 1-800-495-2919 to organize product returns.
Quantity in commerce
90
Distribution
U.S. States: NE, MA, GA, TX, OR, AZ, FL, OH, WA, NJ, UT, VA, CA, MN, PA, DE, OK, KS, WI, MT, IA, MO
Product codes and lots
Lot: 603600

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2339-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2338-2018Class IITerminatedVoluntary: Firm initiated
Z-2339-2018Class IITerminatedVoluntary: Firm initiated