Z-2250-2018Class II

Fabius GS Premium Anesthesia Machine; Cat. no. 8607000

Draeger, Inc. / initiated 2018-05-24 / Terminated

Draeger Medical, Inc. began a recall of Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They... on . FDA classified it as Class II and gives the reason as "The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2250-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Drger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
FDA root cause
Process change control
Action
On May 24, 2018, the firm issued Urgent Medical Device Recall letters to affected end users (medical facilities) via certified mail. The letter identified the affected product, problem and actions to be taken. The recalling firm stated that they are currently manufacturing replacement parts for all potentially affected devices, which will be installed as part of the next service. The firm's Service dept. will contact end users to arrange the replacement Free of Charge In addition, end users were asked to complete and return a Customer Reply Card to acknowledge the receipt of this information. If you have any questions regarding this recall please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349). For questions regarding the operation and/or servicing of your Draeger anesthesia machine please contact DraegerService Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 4 at the prompt then 4 again).
Quantity in commerce
62
Distribution
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Product codes and lots
ASKM-0102 ASKM-0220 ASKM-0103 ASKM-0221 ASKM-0124 ASKM-0222 ASKM-0125 ASKM-0223 ASKM-0129 ASKM-0224 ASKM-0132 ASKM-0229 ASKM-0133 ASKN-0061 ASKM-0134 ASKN-0090 ASKM-0135 ASLA-0035 ASKM-0136 ASLA-0046 ASKM-0137 ASLA-0047 ASKM-0138 ASLA-0048 ASKM-0139 ASLA-0049 ASKM-0189 ASLA-0050 ASKM-0190 ASLA-0053 ASKM-0191 ASLA-0070 ASKM-0192 ASLA-0071 ASKM-0193 ASLA-0072 ASKM-0194 ASLA-0073 ASKM-0195 ASLA-0074 ASKM-0198 ASLA-0075 ASKM-0199 ASLA-0129 ASKM-0200 ASLA-0131 ASKM-0201 ASLA-0133 ASKM-0202 ASLA-0140 ASKM-0204 ASLA-0141 ASKM-0205 ASLA-0159 ASKM-0207 ASLA-0199 ASKM-0208 ASLA-0200 ASKM-0215 ASLA-0201 ASKM-0219 ASLA-0218

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2250-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2250-2018Class IITerminatedVoluntary: Firm initiated
Z-2251-2018Class IITerminatedVoluntary: Firm initiated
Z-2252-2018Class IITerminatedVoluntary: Firm initiated
Z-2253-2018Class IITerminatedVoluntary: Firm initiated
Z-2254-2018Class IITerminatedVoluntary: Firm initiated