510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2278-2018Class II

Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682

Randox Laboratories, Limited / initiated 2018-05-22 / Terminated

Randox Laboratories Ltd. began a recall of Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682 on . FDA classified it as Class II and gives the reason as "The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2278-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.
FDA root cause
Under Investigation by firm
Action
On May 22, 2018, Randox issued an Urgent Field Safety Notice to affected distributors along with an Important Notice and updated IFU. The letter advised customers of the product issue. It further instructed distributors to do the following: 1. Discuss the contents of the notice with your Medical Director. 2. Place a copy of the important notice and updated IFU into any remaining stock. 3. Complete and return the vigilance response section of this form to [email protected] within five working days. If you have any questions after reviewing this information, you may contact the firm at [email protected].
Distribution
Distributed to accounts in CA, IN, KS, ME, and WV, and PR. Foreign distribution to Australia, Bangladesh, Belarus, Belgium, Brazil, Cambodia, Canada, Chile, China, Colombia, Cyprus, Denmark, Dutch Caribbean, Ecuador, Egypt, El Salvador, Ethiopia, Falkland Islands, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Iraq, Ireland, Italy, Jordan, Kenya, Saudi Arabia, Korea, Kuwait, Latvia, Lebanon, Peru, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Morocco, Myanmar, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Oman, Thailand, Netherlands, Turkey, United Arab Emirates, United Kingdom, Ukraine, Vietnam, and Zimbabwe.
Product codes and lots
All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2278-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2278-2018Class IITerminatedVoluntary: Firm initiated
Z-2279-2018Class IITerminatedVoluntary: Firm initiated
Z-2280-2018Class IITerminatedVoluntary: Firm initiated