Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682
Randox Laboratories Ltd. began a recall of Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682 on . FDA classified it as Class II and gives the reason as "The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2278-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.
- FDA root cause
- Under Investigation by firm
- Action
- On May 22, 2018, Randox issued an Urgent Field Safety Notice to affected distributors along with an Important Notice and updated IFU. The letter advised customers of the product issue. It further instructed distributors to do the following: 1. Discuss the contents of the notice with your Medical Director. 2. Place a copy of the important notice and updated IFU into any remaining stock. 3. Complete and return the vigilance response section of this form to [email protected] within five working days. If you have any questions after reviewing this information, you may contact the firm at [email protected].
- Distribution
- Distributed to accounts in CA, IN, KS, ME, and WV, and PR. Foreign distribution to Australia, Bangladesh, Belarus, Belgium, Brazil, Cambodia, Canada, Chile, China, Colombia, Cyprus, Denmark, Dutch Caribbean, Ecuador, Egypt, El Salvador, Ethiopia, Falkland Islands, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Iraq, Ireland, Italy, Jordan, Kenya, Saudi Arabia, Korea, Kuwait, Latvia, Lebanon, Peru, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Morocco, Myanmar, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Oman, Thailand, Netherlands, Turkey, United Arab Emirates, United Kingdom, Ukraine, Vietnam, and Zimbabwe.
- Product codes and lots
- All lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2278-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2279-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2280-2018 | Class II | Terminated | Voluntary: Firm initiated |