510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2284-2018Class II

The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) Proteus 235, Beam...

Ion Beam Applications S.A. / initiated 2016-04-11 / Terminated / root cause: software

Ion Beam Applications S.A. began a recall of The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) Proteus 235, Beam Management System (PBS) on . FDA classified it as Class II and gives the reason as "IBA is conducting a voluntary recall to address a PTS (Proton Therapy System) software issue and to reduce the risk related to this issue". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2284-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ion Beam Applications S.A.
Product
The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) Proteus 235, Beam Management System (PBS)
Product code
LHN System, Radiation Therapy, Charged-Particle, Medical
Reason for recall
IBA is conducting a voluntary recall to address a PTS (Proton Therapy System) software issue and to reduce the risk related to this issue.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
IBA mailed an Urgent Field Safety Notice to affected customers to inform them and prevent risk associated with this issue. IBA also recommends users always check for any discrepancy between the prescribed and delivered MUs on their clinical screen after an irradiation in which a pause occurred. IBA also recommends users always check for any discrepancy between the prescribed and delivered MUs on their clinical screen after an irradiation in which a pause occurred. A software upgrade will be deployed on impacted units to correct this issue.
Quantity in commerce
2
Distribution
Units were distributed to Korea and Jacksonville, FL.
Product codes and lots
PAT.003 and PAT.006

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2284-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2284-2018Class IITerminated