Z-2109-2018Class II

12C (AdaPTinsight) Affected component: AdaPTinsight software

Ion Beam Applications S.A. / initiated 2016-10-28 / Terminated / root cause: software

Ion Beam Applications S.A. began a recall of 12C (AdaPTinsight) Affected component: AdaPTinsight software Product Usage: 12C is used with a charged particle or photon radiation therapy system for... on . FDA classified it as Class II and gives the reason as "IBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2109-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ion Beam Applications S.A.
Product
12C (AdaPTinsight) Affected component: AdaPTinsight software Product Usage: 12C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy system.
Product code
LHN System, Radiation Therapy, Charged-Particle, Medical
Reason for recall
IBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem.
Root cause tag
software
FDA root cause
Software change control
Action
IBA mailed an Urgent Field Safety Notice to affected customers on 10/28/2016 to inform them of the issue. The issue will be solved in AdaPTinsights versions starting from Al 2.0. The solution was to be deployed on impacted IBA Proton Therapy sites by September 2017.
Quantity in commerce
5 units (12C)
Distribution
Worldwide Distribution - US Nationwide in the states of LA, TX, Italy, Sweden and France. Each country's National Competent Authorities were notified of the firm's Field Safety Notice.
Product codes and lots
Affected units: SAT.123, SAT.126, PAT.115, SAT.122 and SBF.101

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2109-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2109-2018Class IITerminatedVoluntary: Firm initiated