Merlin PCS programmer (Model # 3650), Software (Model # 3330)
St Jude Medical Inc. began a recall of Merlin PCS programmer (Model # 3650), Software (Model # 3330) on . FDA classified it as Class II and gives the reason as "The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2329-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Merlin PCS programmer (Model # 3650), Software (Model # 3330)
- Product code
- OSR Pacemaker/Icd/Crt Non-Implanted Components
- Reason for recall
- The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 04/16/2018, Important Medical Device Advisory notifications were mailed via overnight service to U.S. physicians and hospitals notifying customers of the availability of the programmer software update and associated ICD/CRT-D firmware update. Advisory notifications were mailed to customers outside the U.S. on different dates depending on geography and local regulatory requirements. Prophylactic replacement of affected devices is not recommended. The firm recommends firmware upgrades for all eligible patients at the next regularly scheduled visit or when appropriate depending on the preferences of the patient and physician. For devices not eligible for the cybersecurity firmware update the firm recommends a discussion of the risks of cybersecurity vulnerabilities and proven benefits of remote monitoring with patients at their next regularly scheduled visit. RF communication may be permanently disabled in devices not eligible for firmware updates during an in-clinic device interrogation with a programmer that has received the software update. Customers with questions about the firmware update can call the customer technical support hotline at 1-800-436-5056.
- Quantity in commerce
- 71651
- Distribution
- U.S. Nationwide distribution
- Product codes and lots
- Software Versions: 24.1.1 rev1 (US), 23.1.2 rev 1 (OUS), 23.1.3 (Japan), and 23.1.4 rev 1 (Canada)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2318-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2319-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2320-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2321-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2322-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2323-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2324-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2325-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2326-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2327-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2328-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2329-2018 | Class II | Terminated | Voluntary: Firm initiated |