Z-2329-2018Class II

Merlin PCS programmer (Model # 3650), Software (Model # 3330)

Abbott Medical / initiated 2018-04-16 / Terminated / root cause: software

St Jude Medical Inc. began a recall of Merlin PCS programmer (Model # 3650), Software (Model # 3330) on . FDA classified it as Class II and gives the reason as "The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2329-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Merlin PCS programmer (Model # 3650), Software (Model # 3330)
Product code
OSR Pacemaker/Icd/Crt Non-Implanted Components
Reason for recall
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Root cause tag
software
FDA root cause
Software design
Action
On 04/16/2018, Important Medical Device Advisory notifications were mailed via overnight service to U.S. physicians and hospitals notifying customers of the availability of the programmer software update and associated ICD/CRT-D firmware update. Advisory notifications were mailed to customers outside the U.S. on different dates depending on geography and local regulatory requirements. Prophylactic replacement of affected devices is not recommended. The firm recommends firmware upgrades for all eligible patients at the next regularly scheduled visit or when appropriate depending on the preferences of the patient and physician. For devices not eligible for the cybersecurity firmware update the firm recommends a discussion of the risks of cybersecurity vulnerabilities and proven benefits of remote monitoring with patients at their next regularly scheduled visit. RF communication may be permanently disabled in devices not eligible for firmware updates during an in-clinic device interrogation with a programmer that has received the software update. Customers with questions about the firmware update can call the customer technical support hotline at 1-800-436-5056.
Quantity in commerce
71651
Distribution
U.S. Nationwide distribution
Product codes and lots
Software Versions: 24.1.1 rev1 (US), 23.1.2 rev 1 (OUS), 23.1.3 (Japan), and 23.1.4 rev 1 (Canada)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2329-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2318-2018Class IITerminatedVoluntary: Firm initiated
Z-2319-2018Class IITerminatedVoluntary: Firm initiated
Z-2320-2018Class IITerminatedVoluntary: Firm initiated
Z-2321-2018Class IITerminatedVoluntary: Firm initiated
Z-2322-2018Class IITerminatedVoluntary: Firm initiated
Z-2323-2018Class IITerminatedVoluntary: Firm initiated
Z-2324-2018Class IITerminatedVoluntary: Firm initiated
Z-2325-2018Class IITerminatedVoluntary: Firm initiated
Z-2326-2018Class IITerminatedVoluntary: Firm initiated
Z-2327-2018Class IITerminatedVoluntary: Firm initiated
Z-2328-2018Class IITerminatedVoluntary: Firm initiated
Z-2329-2018Class IITerminatedVoluntary: Firm initiated