Ion Beam Applications S.A. began a recall of 12C on . FDA classified it as Class II and gives the reason as "IBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2138-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ion Beam Applications S.A.
- Product
- 12C
- Product code
- LHN System, Radiation Therapy, Charged-Particle, Medical
- Reason for recall
- IBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk.
- FDA root cause
- Finished device change control
- Action
- IBA mailed an Urgent Field Safety Notice to affected customers on 09/16/2016 to notify them of the risks associated with the use of this device to prevent serious injury/death. Customers were asked not to make any unauthorized modification to the IBA System or IBA accessories unless approved in writing or in advance by IBA. IBA will be updating their manuals to include these warnings.
- Quantity in commerce
- S0 Units
- Distribution
- Units were delivered in the USA, Korea, Germany, France, Czech Republic, Italy, Poland and Sweden.
- Product codes and lots
- Code/Serial Numbers: PAT.000, PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.112, PAT.113, PAT.114, PAT.115, SAT.116, SAT.117, SAT.119, SAT.120, SAT.122, SAT.123, SAT.126, SBF.101
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2138-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2139-2018 | Class II | Terminated | Voluntary: Firm initiated |