Z-2343-2018Class II

26-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 2112550

B Braun Medical Inc / initiated 2018-06-04 / Terminated

B. Braun Medical, Inc. began a recall of 26-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 2112550 on . FDA classified it as Class II and gives the reason as "There is a potential for valves on the 26-Lead Transfer Set to be incorrectly oriented. When a transfer set with an incorrectly oriented valve is installed on the system, there is potential for unintended admixture". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2343-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
26-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 2112550
Product code
NEP System/Device, Pharmacy Compounding
Reason for recall
There is a potential for valves on the 26-Lead Transfer Set to be incorrectly oriented. When a transfer set with an incorrectly oriented valve is installed on the system, there is potential for unintended admixture.
FDA root cause
Under Investigation by firm
Action
Urgent Medical Device Correction - Recall Notification letters were distributed on 6/4/18. The letters instruct customers to perform the following: 1. Review this Medical Device Correction Notification in its entirety and ensure that all users of the above mentioned products in your organization and other concerned persons are informed about this medical device correction - recall notification. If you are a distributor, please forward this recall notification to your customers. 2. Perform a visual inspection of valve orientation prior to installation onto APEX Compounding System in accordance with the Visual Inspection Instructions contained in Attachment 1. 3. If any misaligned valves are identified, do not use the transfer set. Please contact Customer Support at 1-800-227-2862 for product return and replacement. 4. Utilizing the attached "Medical Device Correction - Recall Notification Acknowledgement" form, please acknowledge that you have received and reviewed this information. 5. Return the completed "Medical Device Correction - Recall Notification Acknowledgement" form to B. Braun Medical Inc Quality Assurance department by faxing or e-mailing the form within two (2) weeks of receipt of this notice. It is important this form is returned to B. Braun Medical Inc can meet regulatory requirements of the United States Food and Drug Administration.
Quantity in commerce
12320
Distribution
The products were distributed to the following US states: AL, CA, CO, CT, FL, GA, IL, LA, MA, MD, MI, MN, MO, NC, NJ, NV, OH, PA, TX, and WA.
Product codes and lots
0061546906,  0061546910,  0061548145,  0061548146,  0061548476,  0061548477,  0061548478,  0061548479,  0061548645,  0061555100,  0061555101,  0061555102,  0061556350,  0061557521,  0061559077,  0061559628,  0061559631,  0061559852,  0061559853,  0061559854,  0061560830,  0061560831,  0061562493,  0061562494,  0061562941,  0061563541,  0061564175,  0061564176,  0061564177,  0061564179,  0061564180,  0061564431,  0061564669,  0061564670,  0061564672,  0061564673,  0061566362,  0061566363,  0061567272,  0061567809,  0061567810,  0061567811,  0061567812,  0061567813,  0061567814,  0061571324,  0061571325,  0061571326,  0061571327,  0061572464,  0061572465,  0061572466,  0061572980,  0061572981,  0061572982,  0061572983,  0061572984,  0061572985,  0061572987,  0061579647,  0061579648,  0061579651,  0061579653,  0061579654,  0061581522,  0061581585,  0061581586,  0061581590,  0061581599,  0061581602,  0061581603,  0061581982,  0061581983,  0061581984,  0061581985,  0061581986,  0061581987,  0061581988,  0061583281,  0061583282,  0061583283,  0061586220,  0061586221,  0061586222,  0061586223,  0061586224,  0061586712,  0061586713,  0061586714,  0061587151,  0061587152,  0061587153,  0061587154,  0061587155,  0061587156,  0061587157,  0061587720,  0061590070,  0061591603,  0061591618,  0061591646,  0061591660,  0061591662,  0061591663,  0061591664,  0061591665,  0061591667,  0061591675,  0061594365,  0061594366,  0061594367,  0061594368,  0061594369,  0061599040,  0061599159,  0061599160,  0061599161, 0061599162,  0061599163,  0061599164,  0061599165,  0061603919, 0061603920,  0061603921,  0061608039,  0061614625,  0061614626,  00VL559752

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2343-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2343-2018Class IITerminatedVoluntary: Firm initiated