510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2979-2018Class II

Cardiac Insulation Pad, OPEN HEART CDS

MEDLINE INDUSTRIES, LP - Northfield / initiated 2017-04-12 / Terminated / root cause: sterility

Medline Industries, Inc. began a recall of Cardiac Insulation Pad, OPEN HEART CDS on . FDA classified it as Class II and gives the reason as "Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-2979-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Cardiac Insulation Pad, OPEN HEART CDS
Product code
N/A
Reason for recall
Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the CDS assembly. This component should have been placed inside the sterile portion of the pack where it would have subsequently been exposed to an ethylene oxide sterilization process. This component was not placed in the sterile portion of the pack and was not sterilized, resulting in the inadvertent release of this item in a non-sterile state. The contents of the sterile kit were properly sterilized and the remaining contents of the non-sterile subassembly had the appropriate sterility status based on their intended use.
Root cause tag
sterility
FDA root cause
Under Investigation by firm
Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Quantity in commerce
77 kits
Distribution
CA, NM
Product codes and lots
Model: CDS984745C; Lots: 17GD0936, 17FD1655, 17ED3419, 17ED1482, 17ED0636, 17DD0666, 17DD0296, 17AD1912, 16LD3005, 16LD0059, 16JD3201, 16ID3179, 16HD1298, 16GD0779, 16FD3925, 16CD0303 and 16AD4032

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2979-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2976-2018Class IITerminatedVoluntary: Firm initiated
Z-2977-2018Class IITerminatedVoluntary: Firm initiated
Z-2978-2018Class IITerminatedVoluntary: Firm initiated
Z-2979-2018Class IITerminatedVoluntary: Firm initiated
Z-2980-2018Class IITerminatedVoluntary: Firm initiated