Cardiac Insulation Pad, OPEN HEART CDS
Medline Industries, Inc. began a recall of Cardiac Insulation Pad, OPEN HEART CDS on . FDA classified it as Class II and gives the reason as "Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-2979-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Cardiac Insulation Pad, OPEN HEART CDS
- Product code
- N/A
- Reason for recall
- Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the CDS assembly. This component should have been placed inside the sterile portion of the pack where it would have subsequently been exposed to an ethylene oxide sterilization process. This component was not placed in the sterile portion of the pack and was not sterilized, resulting in the inadvertent release of this item in a non-sterile state. The contents of the sterile kit were properly sterilized and the remaining contents of the non-sterile subassembly had the appropriate sterility status based on their intended use.
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
- Quantity in commerce
- 77 kits
- Distribution
- CA, NM
- Product codes and lots
- Model: CDS984745C; Lots: 17GD0936, 17FD1655, 17ED3419, 17ED1482, 17ED0636, 17DD0666, 17DD0296, 17AD1912, 16LD3005, 16LD0059, 16JD3201, 16ID3179, 16HD1298, 16GD0779, 16FD3925, 16CD0303 and 16AD4032
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2976-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2977-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2978-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2979-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2980-2018 | Class II | Terminated | Voluntary: Firm initiated |