Z-2980-2018Class II

Laser Cds

MEDLINE INDUSTRIES, LP - Northfield / initiated 2017-04-12 / Terminated / root cause: labeling and UDI

Medline Industries, Inc. began a recall of LASER CDS on . FDA classified it as Class II and gives the reason as "Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2980-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Laser Cds
Product code
N/A
Reason for recall
Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Quantity in commerce
144 kits
Distribution
CA, NM
Product codes and lots
Model: CDS980970D; Lots: 16FD2090, 16GD2340, 16ID0038, 16KD1940, 17BD1440, 17CD2208

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2980-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2976-2018Class IITerminatedVoluntary: Firm initiated
Z-2977-2018Class IITerminatedVoluntary: Firm initiated
Z-2978-2018Class IITerminatedVoluntary: Firm initiated
Z-2979-2018Class IITerminatedVoluntary: Firm initiated
Z-2980-2018Class IITerminatedVoluntary: Firm initiated