Z-2283-2018Class II

ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000

Siemens Medical Solutions USA, Inc. / initiated 2018-05-01 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000 on . FDA classified it as Class II and gives the reason as "If the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 )...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2283-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
If the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 ). Under certain circumstances, the cover of the detector or collimator can touch and possibly injure the patient due to the C-arm drop. This could lead to a situation in which it is necessary to stop the clinical treatment or continue the treatment on an alternative system.
Root cause tag
design
FDA root cause
Device Design
Action
On May 1, 2018, the firm distributed Urgent Medical Device Correction letters to its affected customers. The letter instructed customers to follow standard emergency procedures they have in place in the event of system failures. The firm recommended considering reinforcing the emergency procedures until the update has been completed. As an immediate measure, operators were instructed to avoid driving the C-arm into a fully extended longitudinal position. The firm is currently developing a solution to eliminate the root cause of the problem. The solution will be distributed to all affected customers upon completion. Customers were asked to forward the safety information to any other organizations that could be affected.
Quantity in commerce
39
Distribution
Healthcare facilities in AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, NE, NJ, NY, OH, TN, TX, UT, and VA. Foreign distribution to Australia, Austria, Belgium, Denmark, France, Germany, Italy, Japan, Kuwait, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
164033 164071 164034 164095 164057 164055 164054 164080 164072 164088 164048 164094 164068 164019 164013 164073 164027 164026 164025 164060 164083 164021 164066 164075 164084 164086 164074 164098 164028 164053 164063 164064 164024 164049 164022 164018 164100 164097 164011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2283-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2283-2018Class IICompletedVoluntary: Firm initiated