Z-2530-2018Class III
Vivo 65, Continuous Ventilator, Home Use, Cat. No. 204000
Breas Medical AB / initiated 2018-05-08 / Terminated / root cause: manufacturing process
Human Design Medical Llc began a recall of Vivo 65, Continuous Ventilator, Home Use, Cat. No. 204000 Product Usage: To provide continuous or intermittent ventilatory support for the care of individuals... on . FDA classified it as Class III and gives the reason as "Some Vivo 65 devices have an unreleased version of the Firmware upgrade tool". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2530-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Breas Medical AB
- Product
- Vivo 65, Continuous Ventilator, Home Use, Cat. No. 204000 Product Usage: To provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation
- Product code
- NOU Continuous, Ventilator, Home Use
- Reason for recall
- Some Vivo 65 devices have an unreleased version of the Firmware upgrade tool.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On May 8, 2018, the firm notified all affected customers via phone to quarantine all affected devices. On May 18, 2018, formal Field Safety Corrective Action letters were sent to all affected customers via FedEx. Customers were asked to remove affected units from use and return to the firm, and to complete and return the Acknowledgement and Product Replacement forms. Customers were informed that the firm would be shipping replacement devices.
- Quantity in commerce
- 45
- Distribution
- US Nationwide Distribution in the states of CA, FL, GA, MI, SC, and TX.
- Product codes and lots
- Pending
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2530-2018 | Class III | Terminated | Voluntary: Firm initiated |