Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2465-2018Class II

PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027

Ethicon, Inc / initiated 2018-05-07 / Terminated / root cause: packaging

Ethicon, Inc. began a recall of PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027 on . FDA classified it as Class II and gives the reason as "The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2465-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
Product code
GAW Suture, Nonabsorbable, Synthetic, Polypropylene
Reason for recall
The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
On May 24, 2018, the firm distributed Urgent: Medical Device Recall letters to affected U.S. customers. Customers were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. 4. Complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-888-473-8015 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall (removal). 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return all unused D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture products that are in their inventory immediately. Only D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture product subject to this recall (removal) returned by September 21, 2018 will be eligible for credit reimbursement. Any product subject to this recall (removal) returned after September 21, 2018 will not be eligible for credit reimbursement. 7. To return product subject to this recall (removal), photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for
Quantity in commerce
144
Distribution
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Product codes and lots
KJQ311 KKE670

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2465-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2448-2018Class IITerminatedVoluntary: Firm initiated
Z-2449-2018Class IITerminatedVoluntary: Firm initiated
Z-2450-2018Class IITerminatedVoluntary: Firm initiated
Z-2451-2018Class IITerminatedVoluntary: Firm initiated
Z-2452-2018Class IITerminatedVoluntary: Firm initiated
Z-2453-2018Class IITerminatedVoluntary: Firm initiated
Z-2454-2018Class IITerminatedVoluntary: Firm initiated
Z-2455-2018Class IITerminatedVoluntary: Firm initiated
Z-2456-2018Class IITerminatedVoluntary: Firm initiated
Z-2457-2018Class IITerminatedVoluntary: Firm initiated
Z-2458-2018Class IITerminatedVoluntary: Firm initiated
Z-2459-2018Class IITerminatedVoluntary: Firm initiated
Z-2460-2018Class IITerminatedVoluntary: Firm initiated
Z-2461-2018Class IITerminatedVoluntary: Firm initiated
Z-2462-2018Class IITerminatedVoluntary: Firm initiated
Z-2463-2018Class IITerminatedVoluntary: Firm initiated
Z-2464-2018Class IITerminatedVoluntary: Firm initiated
Z-2465-2018Class IITerminatedVoluntary: Firm initiated
Z-2466-2018Class IITerminatedVoluntary: Firm initiated