Z-2887-2018Class II

CAREstream ProNox Nitrous Oxide Delivery System labeled as the following: a. CMI-0100 PNX, Pro-Nox Mixer...

Carestream Medical / initiated 2018-05-15 / Terminated

CAREstream Medical LLC began a recall of CAREstream ProNox Nitrous Oxide Delivery System labeled as the following: a. CMI-0100 PNX, Pro-Nox Mixer - Canadian; b. CMI-0100-PNX-HS, Pro-Nox Complete... on . FDA classified it as Class II and gives the reason as "Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine testing". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2887-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Carestream Medical
Product
CAREstream ProNox Nitrous Oxide Delivery System labeled as the following: a. CMI-0100 PNX, Pro-Nox Mixer - Canadian; b. CMI-0100-PNX-HS, Pro-Nox Complete Hospital Roll Stand System - Nitrous Oxide Delivery System; c. CMI-0100-PNX-HS-NO2H, Pro-Nox Hospital System without Oxygen Regulator; d. CMI-0100-PNX-HSWM, Pro-Nox Hospital System Wall Mount with Universal Mount (GCX) scavenger; e. CMI-0100-PNX-NHS, Pro-Nox Complete Roll Stand Nitrous Oxide Delivery System (Non-hospital); f. CMI-0100-PNX-NHS-PS, Pro-Nox Complete Roll Stand Nitrous Oxide Delivery System with Passive Scavenger; g. CMI-0100-PNX-US, Pro-Nox mixer, 6' high-pressure O2 and N2O hoses
Product code
BZR Mixer, Breathing Gases, Anesthesia Inhalation
Reason for recall
Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine testing.
FDA root cause
Under Investigation by firm
Action
CAREstream Medical Ltd. notified its customers on or about 05/15/2015 via a Medical Device Product Correction letter. The letter informed customers to return their units, (model number: CMI-0100-PNX-xxx) for inspection and upgrade at the firms service center in Florida. So as to allow the customer's continued use of the devices safely during this process (and until all units are upgraded), CAREsteam recommends the device only be used with an oxygen monitor with alarms, attached to the device in accordance with our product manual (refer to Users Manual (revision 6) for instructions). Customers were also instructed to fax or email CAREstream Medical Ltd. to acknowledge receipt of the Medical Device Product Correction letter.
Quantity in commerce
919 Units Total
Distribution
Domestic Distribution: AK, AL, AR ,AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. International Distribution: Canada and Morocco.
Product codes and lots
Item Number CMI-0100-PNX-xxx

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2887-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2887-2018Class IITerminatedVoluntary: Firm initiated