Z-2418-2018Class II

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies...

Medtronic, Inc. / initiated 2018-06-21 / Terminated / root cause: design

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage... on . FDA classified it as Class II and gives the reason as "Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2418-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
Root cause tag
design
FDA root cause
Device Design
Action
On 21-Jun-2018, Medtronic began notifying US follow-up physicians, and Risk Managers of health care facilities of this issue via hand delivery by Field Representatives. In consultation with an Independent Physician Quality Panel, MDT recommends the physicians following patients with affected devices: " Consider scheduling an in-office patient follow-up as soon as possible to assess the potential for this issue per the steps described below. " Use a programmer to ensure the Excessive Charge Time EOL&and the Low Battery Voltage ERI& Patient Alerts have been programmed to On-High " Instruct patients to contact your office if they hear device alert tones. " If this issue has occurred, an EOL: replace device immediately Observation will be displayed on the programmer QuickLook report. Schedule device replacement immediately. The notification provided Patient Management recommendations to ensure patient safety and effective high voltage therapy remain as the device battery voltage approaches its 2.61V ERI threshold.
Quantity in commerce
3 units
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
Product UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2418-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2416-2018Class IITerminatedVoluntary: Firm initiated
Z-2417-2018Class IITerminatedVoluntary: Firm initiated
Z-2418-2018Class IITerminatedVoluntary: Firm initiated
Z-2419-2018Class IITerminatedVoluntary: Firm initiated
Z-2420-2018Class IITerminatedVoluntary: Firm initiated
Z-2421-2018Class IITerminatedVoluntary: Firm initiated