Z-2598-2018Class III

BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx...

Becton, Dickinson and Company, BD Biosciences / initiated 2018-05-16 / Terminated / root cause: manufacturing process

Becton, Dickinson and Company, BD Biosciences began a recall of BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton... on . FDA classified it as Class III and gives the reason as "The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2598-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018. If you require further assistance, contact BD Customer/Technical Support BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD representative or distributor.
Quantity in commerce
8 devices
Distribution
US Distribution was made to FL, KS, KY, MA, NJ, and SD.
Product codes and lots
Serial #2662879000001, UDI (01)00382906628794(11)170915(21)Z662879000001;   Serial #2662879000002, UDI (01)00382906628794(11)170915(21)Z662879000002;   Serial #2662879000003, UDI (01)00382906628794(11)170915(21)Z662879000003;   Serial #2662879000004, UDI (01)00382906628794(11)170915(21)Z662879000004;   Serial #2662879000009, UDI (01)00382906628794(11)170915(21)Z662879000009;   Serial #2662879000010; UDI (01)00382906628794(11 )170915(21)Z662879000010;   Serial #2662879000011; UDI (01)00382906628794(11)170915(21)Z662879000011; and   Serial #2662879000012, UDI (01)00382906628794(11)170915(21)Z662879000012.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2598-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2598-2018Class IIITerminatedVoluntary: Firm initiated