510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2435-2018Class II

Edwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm

Edwards Lifesciences LLC / initiated 2018-06-11 / Terminated / root cause: labeling and UDI

Edwards Lifesciences, LLC began a recall of Edwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm on . FDA classified it as Class II and gives the reason as "SAPIEN 3 THV was packaged with the incorrect valve size with respect to the label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2435-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm
Product code
NPT Aortic Valve, Prosthesis, Percutaneously Delivered
Reason for recall
SAPIEN 3 THV was packaged with the incorrect valve size with respect to the label.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
On 06/11/2018, the firm sent an Urgent Product Recall notice, via sales representative hand delivery, to the customer requesting they return the customer response form and the affected device. The customer was advised to call Customer Service at 1-800-424-3278, if they had any additional questions.
Quantity in commerce
1 Valve
Distribution
NJ
Product codes and lots
Use By: 2020-01-24, Serial #: 6005731, UDI #: (01)00690103194357(17)200124(21)6005731

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2435-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2435-2018Class IITerminatedVoluntary: Firm initiated